Nedbrytningsprodukter av nervegasser analysert med væskekromatograf tilkoblet atmospheric pressure ionization elektrospray massespektrometer (HPLC-API-ES-MS) - del II: Negativ ion modus

Date Issued
2002
Keywords
Nervegasser
Massespektroskopi
Project number
2002/01729
Permalink
http://hdl.handle.net/20.500.12242/1535
Collection
Rapporter
02-01729.pdf
Size: 721k
Abstract
Hydrolysis products of nerve agents have been analysed by HPLC coupled to an API-ES-MS in SIM mode. Parameters for the instrument were optimised for MPA, EPA, EMPA, IPMPA and PMPA. The instrument's detection limits for pure standards were 5 ng for the analytes, based on the condition that [M-H]- should be detected with a S/N > 3. The injection volume was 10 μl.